Beckman Coulter Inc. : Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU…
Beckman Coulter DxC 500i Clinical Analyzer Modules and DxC 500 AU Clinical Chemistry Analyzers using software version 1.5 or lower are affected. A software defect can silently overwrite user-customized calculated test rules, potentially generating erroneous patient test results without error notification. The recall is voluntary and firm-initiated.